New Simulation Tool Verifies Doses for Ultra-Compact Upright Proton Therapy
Researchers successfully tested a computer simulation platform to independently check radiation doses for patients treated on the world’s first gantry-less upright proton machine.
Patients receiving radiation therapy on the world’s first ultra-compact upright proton machine can now have their treatment plans independently verified by advanced computer simulations, according to a recent study. Researchers developed and tested a secondary dose-checking system for the Mevion S250-FIT, ensuring that tumor-destroying radiation beams match physician prescriptions with high precision before treatment begins.
Traditional proton therapy requires massive, rotating structures called gantries to aim beams from multiple angles around a lying patient, driving up facility costs and spatial footprints. The S250-FIT system solves this challenge by keeping the proton beam fixed and instead rotating the seated patient on a robotic chair. Because this compact layout features a novel energy-modulation system and smaller beam spots than older models, medical physicists needed an independent method to cross-check the commercial treatment planning software.
To build this independent verification tool, led by researcher Serdar Charyyev and colleagues at Stanford University School of Medicine, the team used a software framework called TOPAS (TOol for PArticle Simulation), which relies on a computational method known as Monte Carlo simulation to track the random paths of millions of subatomic particles. The investigators modeled every physical detail of the machine’s delivery nozzle, including the energy-modulating plates made of boron carbide and polycarbonate, as well as the dynamic adaptive aperture that shapes the beam. They calibrated this virtual nozzle against commissioning measurements taken across 12 different energy settings, ranging from roughly 49 million to 227 million electron volts.

The study results, published on the arXiv preprint server, demonstrate that the simulated proton range agreed with physical measurements within 0.25 millimeters. Furthermore, the spot width matched within 6.8 percent, and the dose per monitor unit agreed within 1.3 percent.
How Well Did the Virtual Simulations Match Real Patient Plans?
To validate the software in practical scenarios, the research team tested an in-house converter program called RayStation2TOPAS on 15 water phantom plans and 25 clinical plans from the first cohort of patients treated with the upright system at Stanford. The clinical cases covered various treatment sites, including 10 brain tumors, 5 head and neck cancers, 6 thoracic tumors, 2 spine cases, and 2 bone cancer cases.
When comparing point doses, the simulation results differed from the commercial treatment planning software by negative 0.76 percent to positive 1.98 percent for the water phantom tests, and by negative 3.03 percent to positive 3.99 percent for the complex clinical plans. Every single plan successfully met the passing criteria for both point dose accuracy and three-dimensional spatial dose distribution analysis, known as gamma analysis.

What Are the Next Steps and Limitations for Clinical Safety?
While the computer simulations successfully passed all validation checks, the authors noted a few operational hurdles for everyday clinical use. Running a complete simulation for a single clinical plan took between 55 and 249 minutes on a multi-core workstation, largely due to the time required to translate complex patient computed tomography scans and structure sets into the simulation environment.
However, because the Stanford facility treats one or two new patients per week on this specific upright platform, medical staff have ample time to run the verification check before a patient’s first fraction of radiation therapy. Future work will focus on streamlining plan preparation to speed up calculation times and further broaden safety protocols for compact proton therapy.
The study was posted to the arXiv preprint server on November 25, 2026.
This article has been fact checked for accuracy, with information verified against reputable sources. Learn more about us and our editorial process.
Last reviewed on .
Article history
- Latest version
Reference(s)
- Alakbarzade, Toghrul. “Monte Carlo-based secondary dose check platform for the world's first ultra-compact upright proton therapy system.” arXiv.org <http://arxiv.org/abs/2611.33924v1>.
Cite this page:
- Posted by Sana Malik