New Simulation Tool Verifies Doses for Ultra-Compact Upright Proton Therapy
Medicine

New Simulation Tool Verifies Doses for Ultra-Compact Upright Proton Therapy

Researchers successfully tested a computer simulation platform to independently check radiation doses for patients treated on the world’s first gantry-less upright proton machine.

By Sana Malik
Published:
Email this Article
New Simulation Tool Verifies Doses for Ultra-Compact Upright Proton Therapy
Photo by National Cancer Institute on Unsplash

Patients receiving radiation therapy on the world’s first ultra-compact upright proton machine can now have their treatment plans independently verified by advanced computer simulations, according to a recent study. Researchers developed and tested a secondary dose-checking system for the Mevion S250-FIT, ensuring that tumor-destroying radiation beams match physician prescriptions with high precision before treatment begins.

Traditional proton therapy requires massive, rotating structures called gantries to aim beams from multiple angles around a lying patient, driving up facility costs and spatial footprints. The S250-FIT system solves this challenge by keeping the proton beam fixed and instead rotating the seated patient on a robotic chair. Because this compact layout features a novel energy-modulation system and smaller beam spots than older models, medical physicists needed an independent method to cross-check the commercial treatment planning software.

To build this independent verification tool, led by researcher Serdar Charyyev and colleagues at Stanford University School of Medicine, the team used a software framework called TOPAS (TOol for PArticle Simulation), which relies on a computational method known as Monte Carlo simulation to track the random paths of millions of subatomic particles. The investigators modeled every physical detail of the machine’s delivery nozzle, including the energy-modulating plates made of boron carbide and polycarbonate, as well as the dynamic adaptive aperture that shapes the beam. They calibrated this virtual nozzle against commissioning measurements taken across 12 different energy settings, ranging from roughly 49 million to 227 million electron volts.

Figure 1: The TOPAS model of the treatment nozzle, drawn by TOPAS from the input the converter wrote for one energy layer of the static plan of the 1000 cm31000\text{\,}{\mathrm{cm}}^{3} target at 10 cm10\text{\,}\mathrm{c}\mathrm{m} depth (section 2.5), with the snout at 200 mm200\text{\,}\mathrm{m}\mathrm{m}. The arrow shows the direction of the beam, and the lines point from the labels to the components. The EMS plates (red) in this layer are in the beam, while the other plates are retracted, all inside the EMS air box. The AA (yellow) is a nickel cylinder with the opening of the plan (gray), followed by the exit window (green) and the water phantom (blue), whose surface is 50 mm50\text{\,}\mathrm{m}\mathrm{m} upstream of the isocenter (cross). The virtual source, 1870 mm1870\text{\,}\mathrm{m}\mathrm{m} upstream of the isocenter, is outside the view.
The TOPAS model of the treatment nozzle, drawn by TOPAS from the input the converter wrote for one energy layer of the static plan of the 1000 cm31000\text{\,}{\mathrm{cm}}^{3} target at 10 cm10\text{\,}\mathrm{c}\mathrm{m} depth (section 2.5), with the snout at 200 mm200\text{\,}\mathrm{m}\mathrm{m}. The arrow shows the direction of the beam, and the lines point from the labels to the components. The EMS plates (red) in this layer are in the beam, while the other plates are retracted, all inside the EMS air box. The AA (yellow) is a nickel cylinder with the opening of the plan (gray), followed by the exit window (green) and the water phantom (blue), whose surface is 50 mm50\text{\,}\mathrm{m}\mathrm{m} upstream of the isocenter (cross). The virtual source, 1870 mm1870\text{\,}\mathrm{m}\mathrm{m} upstream of the isocenter, is outside the view. Credit: Toghrul Alakbarzade et al. / arXiv

The study results, published on the arXiv preprint server, demonstrate that the simulated proton range agreed with physical measurements within 0.25 millimeters. Furthermore, the spot width matched within 6.8 percent, and the dose per monitor unit agreed within 1.3 percent.

How Well Did the Virtual Simulations Match Real Patient Plans?

To validate the software in practical scenarios, the research team tested an in-house converter program called RayStation2TOPAS on 15 water phantom plans and 25 clinical plans from the first cohort of patients treated with the upright system at Stanford. The clinical cases covered various treatment sites, including 10 brain tumors, 5 head and neck cancers, 6 thoracic tumors, 2 spine cases, and 2 bone cancer cases.

When comparing point doses, the simulation results differed from the commercial treatment planning software by negative 0.76 percent to positive 1.98 percent for the water phantom tests, and by negative 3.03 percent to positive 3.99 percent for the complex clinical plans. Every single plan successfully met the passing criteria for both point dose accuracy and three-dimensional spatial dose distribution analysis, known as gamma analysis.

Figure 2: The patient pipeline, shown on the CT of the IROC Houston anthropomorphic head-and-neck phantom, for one beam of its plan, on the axial slice through the isocenter. (a) Planning CT with the External outline and the box to which the CT was cropped. (b) The CT as TOPAS reads it, with the voxels outside the External set to air, the direction of the beam, and the snout face of the nozzle. (c) RayStation and (d) TOPAS dose of the beam, as physical dose, with TOPAS resampled onto the RayStation grid and set to zero outside the External. Dose below 10 % of the maximum is not shown. (e) Gamma index at 3 %/2 mm. Because a single beam is shown, the dose-difference criterion and the low-dose threshold were percentages of the maximum of the RayStation dose of the beam, not of the dose at the isocenter as for a plan. The cross marks the isocenter.
The patient pipeline, shown on the CT of the IROC Houston anthropomorphic head-and-neck phantom, for one beam of its plan, on the axial slice through the isocenter. (a) Planning CT with the External outline and the box to which the CT was cropped. (b) The CT as TOPAS reads it, with the voxels outside the External set to air, the direction of the beam, and the snout face of the nozzle. (c) RayStation and (d) TOPAS dose of the beam, as physical dose, with TOPAS resampled onto the RayStation grid and set to zero outside the External. Dose below 10 % of the maximum is not shown. (e) Gamma index at 3 %/2 mm. Because a single beam is shown, the dose-difference criterion and the low-dose threshold were percentages of the maximum of the RayStation dose of the beam, not of the dose at the isocenter as for a plan. The cross marks the isocenter. Credit: Toghrul Alakbarzade et al. / arXiv

What Are the Next Steps and Limitations for Clinical Safety?

While the computer simulations successfully passed all validation checks, the authors noted a few operational hurdles for everyday clinical use. Running a complete simulation for a single clinical plan took between 55 and 249 minutes on a multi-core workstation, largely due to the time required to translate complex patient computed tomography scans and structure sets into the simulation environment.

However, because the Stanford facility treats one or two new patients per week on this specific upright platform, medical staff have ample time to run the verification check before a patient’s first fraction of radiation therapy. Future work will focus on streamlining plan preparation to speed up calculation times and further broaden safety protocols for compact proton therapy.

The study was posted to the arXiv preprint server on November 25, 2026.

Fact Checked

This article has been fact checked for accuracy, with information verified against reputable sources. Learn more about us and our editorial process.

Last reviewed on .

Article history

  • Latest version

Reference(s)

  1. Alakbarzade, Toghrul. “Monte Carlo-based secondary dose check platform for the world's first ultra-compact upright proton therapy system.” arXiv.org <http://arxiv.org/abs/2611.33924v1>.

Cite this page:

Malik, Sana. “New Simulation Tool Verifies Doses for Ultra-Compact Upright Proton Therapy.” BioScience. BioScience ISSN 2521-5760, 26 November 2026. <https://www.bioscience.com.pk/en/subject/medicine/monte-carlo-based-secondary-dose-check-platform-for-the-worlds-first-ultra-compact-upright-proton-therapy-system>. Malik, S. (2026, November 26). “New Simulation Tool Verifies Doses for Ultra-Compact Upright Proton Therapy.” BioScience. ISSN 2521-5760. Retrieved November 26, 2026 from https://www.bioscience.com.pk/en/subject/medicine/monte-carlo-based-secondary-dose-check-platform-for-the-worlds-first-ultra-compact-upright-proton-therapy-system Malik, Sana. “New Simulation Tool Verifies Doses for Ultra-Compact Upright Proton Therapy.” BioScience. ISSN 2521-5760. https://www.bioscience.com.pk/en/subject/medicine/monte-carlo-based-secondary-dose-check-platform-for-the-worlds-first-ultra-compact-upright-proton-therapy-system (accessed November 26, 2026).