Wearable Textile Sensor Tracks Sweat Salinity And Skin Saturation In Real Time
A new cotton patch measures ionic currents during exercise, matching thermal camera and heart rate data without relying on delicate chemical coatings.
A new flexible textile sensor tracks sweat salinity and skin saturation in real time during physical activity without relying on delicate chemical coatings. Researchers developed the wearable patch to monitor physical exertion by measuring electrical currents directly on the skin.
The device consists of cotton fibers treated with a conductive polymer known as PEDOT:PSS, which forms both the channel and gate electrodes of an organic electrochemical transistor. When sweat enters the patch, ions penetrate the polymer bulk and alter the electrical current flowing through the fibers. Francesco Gentile from the University Magna Graecia of Catanzaro and their collaborators published in ACS Omega that this configuration avoids the need for selective enzymes or ion-selective membranes, which are prone to fouling and calibration drift.
How Does A Textile Patch Decode Salt Concentration And Moisture Levels?
Traditional sweat sensors typically target specific individual molecules using specialized chemical receptors. This new system instead measures the total electrical response and applies a physics-based mathematical model to separate overall salt concentration from the degree of fabric saturation. As sweat fills the porous cotton structure, changes in the gate and drain currents are recorded every sixty seconds.

The research team tested the sensor on three adult participants during treadmill running sessions lasting up to one hour. The participants were monitored while running through varying speeds and inclines, with workloads designed to induce gradual physiological stress. By processing the electrical signals through their mathematical framework, the authors were able to estimate sweat salinity ranging from low baseline values up to approximately eighty millimolar, alongside skin saturation levels ranging from zero to one hundred percent.
What Did Body Temperature And Heart Rate Comparisons Reveal?
To validate the sensor output, the team simultaneously recorded thermal infrared videos of the runners’ faces and monitored their heart electrical activity using electrocardiographic chest leads. Comparing the patch data against facial skin temperature across six distinct zones revealed correlation coefficients as high as 0.90. A similarly strong correlation of up to 0.86 was found between sensor-derived saturation and heart rate variability in two of the three participants.
The cross-correlation analysis also uncovered a distinct physiological time lag between internal heating and the appearance of sweat. The delay between the onset of exercise-induced temperature increases and measurable sweat saturation ranged from six hundred to seven hundred and forty seconds across the trials. This temporal gap reflects the natural delay required for thermoregulatory sweat production to activate following increased physical exertion.

What Are The Current Limitations And Technical Constraints?
While the sensor successfully tracked continuous sweat dynamics, the current design involves certain technical compromises. The physical process of ion diffusion into the polymer channel limits the sampling frequency to discrete measurements spaced sixty seconds apart. Furthermore, the mathematical model simplifies all dissolved sweat electrolytes into an equivalent sodium chloride solution. Control experiments involving alternative salts such as potassium chloride and calcium chloride indicated that the electrical response changes by less than seventeen percent, suggesting that the sodium chloride approximation remains a reliable first-order baseline under normal physiological conditions.
The authors note that the investigation serves as a proof-of-concept tested on a small group of three individuals rather than a large clinical cohort. Future developments aim to reduce the dimensions of the conductive channels to speed up response times and to integrate multi-sensor arrays for higher temporal resolution. The platform provides a foundation for continuous noninvasive monitoring that could eventually assist in evaluating hydration disorders, metabolic function, and athletic recovery.
The research was published in ACS omega on November 1, 2026.
This article has been fact checked for accuracy, with information verified against reputable sources. Learn more about us and our editorial process.
Last reviewed on .
Article history
- Latest version
Reference(s)
- Gentile, Francesco., et al. “Wearable Organic Electrochemical Transistor Biosensor for Real-Time Measurement of Sweat Salinity and Saturation.” ACS Omega, vol. 11, no. 45, November 1, 2026, pp. 69801-69817. American Chemical Society (ACS), doi: 10.1021/acsomega.6c05587. <https://doi.org/10.1021/acsomega.6c05587>.